Agilent Technologies 54502A Manual de Serviço Página 64

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 111
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 63
b. Move cursor to bit assignment field and turn on the appropriate eight
bits to be tested (* = on; . = off).
4. Set Timing Trace Specification as in figure 3-37. Follow steps a through c if
unfamiliar with menus.
a. Move cursor to Acquisition mode and SELECT Glitch.
b. Move cursor to Find Pattern and SELECT. Push DONT CARE (all
X s) and SELECT. Set present for field to > 30 ns.
c. Set Then find Glitch to all DONT CARES (all
s).
Figure 3-36. Threshold Accuracy Format Specification
Figure 3-37. Threshold Accuracy Trace Specification
HP 1650B/1651B Performance Tests
Service Manual 3-29
Vista de página 63
1 2 ... 59 60 61 62 63 64 65 66 67 68 69 ... 110 111

Comentários a estes Manuais

Sem comentários