7. Connect the next clock to the test connector and repeat steps 4 and 6 for the
appropriate clock. Repeat until all clocks have been tested (clocks J, K, L,
M and N).
8. Remove the probe tip assembly from the logic analyzer probe cable and
attach to the next logic analyzer probe cable to be tested. Take care not to
dislodge grabbers from the test connector.
9. Repeat steps 3, 4, 6 and 7 until all pods have been tested (pods 1 through 5).
Start again with the rising edge of J clock and test the pod with all clocks.
10. Repeat until the lower bits of all pods have been tested (pods 1 through 5).
11. Disconnect lower eight bits (bits 0 through 7) from test connector. Attach
upper eight bits (bits 8 through 15) to the test connector. Repeat steps 3, 4,
6, 7, and 8 until the upper bits of all pods have been tested (pods 1 through 5).
Performance Tests HP 1650B/1651B
3-12 Service Manual
Comentários a estes Manuais