5. Set the State Trace Specification without sequencing levels and set Count to
Off as in Figure 3-13.
6. Press RUN. The State Listing is displayed and lists all 0s for the channels
under test, if passed test, as in figure 3-14.
Note To ensure consistent pattern of 0s in listing, use front-panel ROLL keys and knob
to scroll through State Listing.
Figure 3-13. State Listing for Data Test 2
Figure 3-14. Setup for Data Test 3
HP 1650B/1651B Performance Tests
Service Manual 3-11
Vista de página 45