Chapter 3: Testing Performance
To Test the Single-clock, Multiple-edge, State Acquisition
3 Set up the Format tab.
a Under one of the pod fields, select TTL.
b In the Pod Threshold window, ensure the Apply threshold setting to all
pods checkbox is checked.
c In the Pod Threshold window, select the Standard threshold voltage
field. At the pop-up menu, select ECL (–1.30 V).
d Select Close to close the Pod Threshold window.