Chapter 5: Troubleshooting
7 On the logic analyzer, select Run. The listing should look similar to the
figure below. Ignore any error messages dealing with the G1 and G2
markers.
8 If the listing looks like the figure, then the cable passed the test.
If the listing does not look similar to the figure, then there is a possible problem
with the cable or probe tip assembly. Causes for cable test failures include:
open channel.
channel shorted to a neighboring channel.
channel shorted to either ground or a supply voltage.
Return to the troubleshooting flowchart.