Agilent 1260 Infinity High Performance Autosampler User Manual 135
Maintenance
9
Removing the Metering Seal
Removing the Metering Seal
When When poor injection volume reproducibility or when metering device / analytical head is leaking.
Tools required p/n Description
8710-0510 Wrench open 1/4 — 5/16 inch
8710-2392 4 mm Hex key
G4226-43800 Seal insert tool
Parts required # p/n Description
1 5063-6589 Metering seal (pack of 2) for 100 µL analytical head
1 In the user interface start the maintenance mode and
select Change metering device function. In the Agilent
Lab Advisor software the Change metering device
function can be found in the Tools section.
2 Open the front door.