128 Agilent 1260 Infinity High Performance Autosampler User Manual
9 Maintenance
Installing the needle assembly
1
Push the silicon safety tube delivered with every needle
over the needle.
2 Insert the loop capillary into the needle assembly and
tighten the fitting hand tight.
3 Pinch the holder clamp and reinsert the needle assembly
into the needle carrier.
4 Attach a 5/16 inch wrench to hold the position at the
needle assembly. Use a 1/4 inch wrench to tighten the
fitting of the loop capillary.