134 Agilent 1260 Infinity High Performance Autosampler User Manual
9 Maintenance
Replacing the Rotor seal
7
Install new rotor seal and isolation seal. 8 Reinstall the stator ring and stator head. The pins on
stator ring and stator head must engage in the
corresponding holes.
9 Insert and tighten the stator screws alternating with the
9/64 inch hex key, until the stator head is secure.
Next Steps:
10 Reconnect all capillaries to the injection valve ports with
a 1/4 inch wrench. The positions of the individual fittings
can be seen on the sticker on the sampling unit.
11 Close the front door.