ESG Family Signal Generators Equipment Required
Required Equipment
Calibration Guide 1-3
• HP/Agilent 8904A Function
Generator
1
Frequency range (sinewave):
0 Hz to 600 kHz
AC Amplitude (sinewave only):
Range: 0 to 10 Vp-p into a 50Ω load
Flatness (>630 mVp-p into 50Ω):
±0.1% (±0.009 dB), 0.1 Hz to 100 kHz
• HP/Agilent 53132A Option 050
Universal Counter
1
Frequency range: 4 GHz
Resolution: 0.1 Hz
External timebase input
• HP/Agilent 33120A Option 001
Arbitrary Waveform Generator
1
(2)
(No Substitution)
• HP/Agilent 438A Power Meter
1
Instrumentation accuracy: ±0.5%
Power reference accuracy: 0.9%
• HP/Agilent 8482A Power Sensor
Frequency range: 100 kHz to 4000 MHz
Power range: −30 dBm to +13 dBm
Maximum SWR:
100 kHz to 300 kHz: 1:1.6,
300 kHz to 1 MHz: 1:1.12,
1 MHz to 2 GHz: 1:1.1,
2 GHz to 4 GHz: 1:1.3
Calibration factor accuracy (RSS):
≤ 1.6%
• HP/Agilent E3615A Power Supply
Minimum Power: 15V
Minimum Current: 325 mA
• MD/MDC-174 Mixer
Frequency range: 0.001 to 2.8 GHz
• MD/MDC-164 Mixer
Frequency range: 0.5 to 9 GHz
• HP/Agilent 3458A Digital
Multimeter
1
dc Volts: 0.1 to 100 V
Input resistance: 10Ω to 1 KΩ
• HP/Agilent 8491A/B Option 006
Attenuator
Attenuation: 6 dB
• HP/Agilent 8491A/B Option 010
Attenuator
Attenuation: 10 dB
• HP/Agilent 8491A/B Option 020
Attenuator
Attenuation: 20 dB
• Personal Computer (included in
HP/Agilent E5500A/B)
Running MS Windows® 3.1/95/98 or
Windows NT®
• HP LaserJet 5L Printer
MS Windows supported
• HP 82341C GPIB Interface Card
with SICL HP I/O Library Installed
(No Substitution)
• Agilent Technologies E4400-60073
Digital Test Card (No Substitution)
• HP/Agilent 8495G Programmable
Step Attenuator
Attenuation: 0 to 70 dB
Frequency range: dc to 4 GHz
• HP/Agilent 11713A Step Attenuator
Driver
• HP/Agilent E5500A/B Phase Noise
Measurement System
2, 3
Frequency range: 0 to 4 GHz
1. Equipped with GPIB
2. Includes Personal Computer
3. Required for ESG-AP/DP only