ESG Family Signal Generators Service Software
ESG Family Support Software Administration
Calibration Guide 3-9
Figure 3-3 Adding Test Equipment Using the Test Equipment Menu
3. Select the model of the device.
4. Click Add.
5. In the New Test Equipment window (refer to Figure 3-4), enter the following
parameters into the appropriate box:
Serial Number of the new equipment.
GPIB Address of the new equipment. This address must be in the range of 0
through 31 and it should not conflict with any other instrument
address already present in the test setup.
NOTE The power sensor must be assigned GPIB address 1 (negative 1).
Cal Due Date the projected calibration due date of the new equipment.
Trace Number the calibration tracking number. This is the last required item.
Calibration Type (optional) is used only when special calibration data is associated
with the device being added, and only as a function setup by the
factory for equipment requiring specific calibration data. The only
devices currently requiring this feature are power sensors.
Table 3-1 is an example of the calibration information required
to ensure accuracy for measurements using the power sensor.
This window is accessed for data entry by selecting the words
CAL DATA from the Calibration Type field in the Edit or New
Test Equipment window, Figure 3-4.