Agilent Technologies E4438C Guia de Instalação Página 116

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 140
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 115
Adjustments
DCFM Calibration
4-24
DCFM Calibration
Description
This calibration removes all of the dc offsets associated with the FM path while in DCFM mode. This test uses
only the FM1 path to verify the functionality of the circuitry. The resulting values are stored in the signal
generator’s firmware.
Required Test Equipment
Agilent 53132A Option 050 frequency counter
Procedure
Figure 4-17 DCFM Calibration Setup
1. Connect the equipment as shown above.
2. Preset all of the equipment.
3. Follow the instructions as they appear on the controllers display.
Vista de página 115
1 2 ... 111 112 113 114 115 116 117 118 119 120 121 ... 139 140

Comentários a estes Manuais

Sem comentários