3-56 Agilent E6432A Microwave Synthesizer Users Guide
Soft Front Panel Help
Configuration Dialog Box
FM Sensitivity (Option 002 Only)
This option is not available on instruments with Option UNG.
FM Sensitivity, available from the pull down View menu by selecting
Configuration, is used to select the FM sensitivity. It can be set to
10 MHz/V, 1 MHz/V, or 100 KHz/V.
Level accuracy with ALC off below 2 GHz is unspecified.
A Power Search may be used to improve level accuracy with ALC off.
Alternately, the IF Attenuator, available from the pull down View menu
by selecting Configuration, can be adjusted in 2 dB steps to obtain the
correct level within +/- 1 dB; above 2 GHz or with ALC on, this is
unnecessary.
If the modulation index, defined as the peak frequency deviation
divided by the frequency of the modulating signal, is greater than ~120,
a frequency shift may occur.
For complete specifications, refer to “Specifications and
Characteristics” on page 6-1.