2-16 Programming Guide
Programming Commands and Examples ESG Family Signal Generators
:DIAGnostic Subsystem
:DIAGnostic Subsystem
The diagnostic subsystem is used to set the controls and the parameters associated with
instrument operational and tracking data.
Attenuator Cycle Information
:DIAGnostic[:CPU]:INFOrmation:CCOunt:ATTenuator?
This query returns the number of times that the attenuator has been switched.
Display Time-On Information
:DIAGnostic[:CPU]:INFOrmation:DISPlay:OTIMe?
This query returns the number of hours that the source’s display has been activated.
Power Cycle Information
:DIAGnostic[:CPU]:INFOrmation:CCOunt:PON?
This query returns the number of times that the sources line power has been cycled.
Reverse Power Protection Trips Information
:DIAGnostic[:CPU]:INFOrmation:CCOunt:PROTection?
This query returns the number of times that the reverse power protection circuitry has
been activated.
Option Information
:DIAGnostic[:CPU]:INFOrmation:OPTions?
This query returns a list of installed instrument options. For instance: UN8, UN7, 300.
Detailed Option Information
:DIAGnostic[:CPU]:INFOrmation:OPTions:DETail?
This query returns information on the installed instrument options.
Installed Board Information
:DIAGnostic[:CPU]:INFOrmation:BOARds?
This query returns a list of the installed boards in the instrument, including the board
name, part number, serial number, version number and status.