Agilent Technologies ESG Especificações Página 46

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 259
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 45
1-42 Programming Guide
Preparing for Use ESG Family Signal Generators
Programming the Status Register System
Programming the Status Register System
The signal generator’s instrument status system provides complete IEEE 488.2 Device
Standard data structures for reporting instrument status using the register model.
The IEEE 488.2 register model of the status system is comprised of multiple registers
which are arranged in a hierarchical order. The lower-priority status registers propagate
their data to the higher-priority registers in the data structures by means of summary bits.
The Status Byte Register is at the top of the hierarchy and contains the general status
information for the instrument’s events and conditions. All other individual registers are
used to determine the specific events or conditions.
You can determine the state of certain instrument hardware and firmware events and
conditions by programming the status register system.