Programming Guide 2-93
ESG Family Signal Generators Programming Commands and Examples
Generating an External DC-Coupled Pulse Modulated Signal, Example Program 8
Program Comments
10 to 160: Title and program description
170: Assigns the signal generator’s GPIB address to a variable.
180: Places the signal generator into LOCAL mode.
190: Resets the signal generator’s parser and clears any pending output from the source.
200: Clears the controller’s display.
210: Sets the signal generator to a defined state for programming.
220: Sets the signal generators pulse modulation source to EXT2.
230: Turns pulse modulation on.
240: Sets the carrier frequency to 5 MHz.
250: Turns off the power leveling search to assist in maintaining the required power level.
260: Sets the output power level to 5 dBm.
270: Turns on the power leveling search to assist in maintaining the required power level.
280: Turns the RF output on.
290 to 300: Prints a message to the controller’s display.
310: Places the signal generator into LOCAL mode.
320: Ends the program.