Programming Guide 1-53
ESG Family Signal Generators Preparing for Use
Programming the Status Register System
The Standard Operation Condition Register continuously monitors the hardware and
firmware status of the instrument. Condition registers are read-only. To query the
condition register, send the command STATus:OPERation:CONDition? The response will
be the decimal sum of the bits which are set to 1. For example, if bit number 9 and bit
number 3 are set to 1, the decimal sum of the 2 bits is 512 plus 8. So the decimal value 520
is returned.
The transition filter specifies which types of bit state changes in the condition register will
set corresponding bits in the event register. The changes may be positive (from 0 to 1) or
negative (from 1 to 0). Send the command STATus:OPERation:NTRansition <num>
(negative) or STATus:OPERation:PTRansition <num> (positive) where <num> is the sum of
the decimal values of the bits you want to enable.
The Standard Operation Event Register latches transition events from the condition
register as specified by the transition filters. Event registers are destructive read-only.
Reading data from an event register will clear the content of that register. To query the
event register, send the command STATus:OPERation[:EVENt]?
Figure 1-25.
The Standard Operation Status Group also contains a Standard Operation Event Enable
Register. This register lets you choose which bits in the Standard Operation Event
Register will set the summary bit (bit 7 of the Status Byte Register) to 1. Send the
STATus:OPERation:ENABle <num> command where <num> is the sum of the decimal
values of the bits you want to enable. For example, to enable bit 9 and bit 3 so that
whenever either of those bits is set to 1, the Standard Operation Status summary bit of the
Status Byte Register will be set to 1, send the command STAT:OPER:ENAB 520 (512 + 8).
The command STATus:OPERation:ENABle? returns the decimal value of the sum of the
bits previously enabled with the STATus:OPERation:ENABle <num> command.