Agilent Technologies E5500 Series Guia do Utilizador Página 142

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 232
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 141
Scanner Maintenance and Calibration 8
Agilent 5500 SPM User’s Guide 142
movement) and a single tube for Z motion. Since the range is small with
these scanners, the effect of cross-coupling is minimal.
Aging
Aging is a time-dependent effect in which the sensitivity (extension per
unit of field) of the scanner decreases, approximately exponentially,
over time. A large amount of decrease takes place during the first few
hours of use. Therefore, scanners are burned in before initial calibration.
After this initial burn-in, the aging rate is very slow; however,
calibration once or twice per year is still recommended.
Creep
At a constant applied voltage the piezo position will change slightly
over time, most noticeably at the beginning of a scan or after a change
on scan location. Creep appears as elongation of the feature in the
direction of the offset because of this slow movement (Figure 93).
Figure 93 Scanner creep
Vista de página 141
1 2 ... 137 138 139 140 141 142 143 144 145 146 147 ... 231 232

Comentários a estes Manuais

Sem comentários