Agilent Technologies E5500 Series Guia do Utilizador Página 116

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 232
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 115
Additional Imaging Modes 7
Agilent 5500 SPM User’s Guide 116
located beneath the tip, can be field-replaced to adjust the sensitivity if
necessary.
Figure 81 STM scanner
The procedure for STM imaging is as follows:
1 If you are using the multi-purpose scanner, insert the nose assembly
into the scanner (see Chapter 4 for details).
2 Insert a tip into the nose assembly or scanner. Grasp the tip with a
tweezers, then insert it into the hollow tube until it protrudes
approximately 2 mm (Figure 82).
Figure 82 Inserting STM tip wire in scanner
3 Place the scanner in the scanner bracket on the microscope.
4 Prepare the sample and place it on a sample plate. The sample must
be electrically isolated from the sample plate. The particular
mounting arrangement will depend on the sample type and size.
Vista de página 115
1 2 ... 111 112 113 114 115 116 117 118 119 120 121 ... 231 232

Comentários a estes Manuais

Sem comentários