Agilent Technologies U2000 Series Manual do Utilizador Página 15

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 24
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 14
15
Specifications (continued)
Settling time
Manual filter, 10-dB decreasing power step (not across switching points)
Number of averages 1 2 4 8 16 32 64 128 256 512 1024
Settling time (s)
Normal speed
x2 speed
0.045
0.042
0.09
0.05
0.17
0.09
0.34
0.17
0.66
0.34
1.3
0.66
2.6
1.3
5.2
2.6
10.4
5.2
20.9
10.4
41.9
20.9
* When a power step crosses through the sensor’s auto-range switching point, add 25 ms
Settling time with auto filter, default resolution, and a 10 dB decreasing power step (not across the switching point)
Settling time = 25 ms*
Auto filter, default resolution, 10-dB decreasing power step
Typical
settling
times
x2
speed
Normal
speed
45 ms 45 ms
82 ms 90 ms
1.3 s 2.6 s
1.5 s 2.7 s
450 ms 460 ms
1.6 s 2.8 s
20 s 39 s
24 s 42 s
24 s 42 s
U2000/1/2/4A U2000/1/2H U2000/1B
+10 dBm
+2 dBm
–4 dBm
–10 dBm
–20 dBm
–30 dBm
–40 dBm
–50 dBm
+20 dBm
+12 dBm
+6 dBm
0 dBm
–10 dBm
–20 dBm
–30 dBm
–40 dBm
+40 dBm
+32 dBm
+26 dBm
+20 dBm
+10 dBm
0 dBm
–10 dBm
–20 dBm
Sensor
dynamic
range
Maximum dBm
Minimum dBm
Low power path
High power path
Vista de página 14
1 2 ... 10 11 12 13 14 15 16 17 18 19 20 ... 23 24

Comentários a estes Manuais

Sem comentários