1 Introduction
The U3606B in Brief
26 U3606B User’s Guide
Table 1 - 6 Output terminal connections
Function Output terminals (+ FORCE –) Maximum output
Constant voltage output
Amplitude: 0 V to 30 V
[1]
OCP
S1: 0 A to 1.1 A
S2: 0 A to 3.3 A
S2m: 0 A to 3.3 A
S1S2: 0 A to 3.3 A
OC:
S1: 0 A to 1.05 A
S2: 0 A to 3.15 A
S2m: 0 A to 3.15 A
S1S2: 0 A to 3.15 A
Constant current output
Amplitude: 0 A to 3 A
[2]
OVP:
S1: 0 V to 33 V
S2: 0 V to 8.8 V
S1m: 0 V to 33 V
S1S2: 0 V to 33 V
OV:
S1: 0 V to 31.5 V
S2: 0 V to 8.4 V
S1m: 0 V to 31.5 V
S1S2: 0 V to 31.5 V
Ramp output
Amplitude:
CV: 0 V to 31.5 V
[1]
CC: 0 A to 3.15 A
[2]
Number of steps: 1 to 10000 steps
Scan output
Amplitude:
CV: 0 V to 31.5 V
[1]
CC: 0 A to 3.15 A
[2]
Number of steps: 1 to 100 steps
Dwelling time: 1 to 99 s