Agilent Technologies U3606B Guia do Utilizador Página 181

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 205
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 180
Characteristics and Specifications 5
Digital Multimeter Specifications
U3606B User’s Guide 165
Frequency specifications
Table 5 - 3 Frequency accuracy specifications ± (% of reading + % of range)
Function Range Frequency range 1 Year
23 ºC ± 5 ºC
Temperature
coefficient
0 ºC to 18 ºC
28 ºC to 55 ºC
Frequency
[1]
Voltage path:
100 mV to 750 V
< 2 Hz 0.18 + 0.003 0.005
< 20 Hz 0.04 + 0.003 0.005
20 Hz to 100 kHz 0.02 + 0.003 0.005
100 kHz to 300 kHz 0.02 + 0.003 0.005
Current path:
10 mA to 3 A
< 2 Hz 0.18 + 0.003 0.005
< 20 Hz 0.04 + 0.003 0.005
20 Hz to 10 kHz 0.02 + 0.003 0.005
[1] For 100 mV and 1 V ranges, the measurable frequency is up to 1 MHz at 0.5 V signal . Minimum input frequency is 1 Hz.
NOTE
All frequency counters are susceptible to errors when measuring low-voltage,
low-frequency signals. Shielding inputs from external noise pickup is critical for
minimizing measurement errors.
U3606-90054_UG_EN.book Page 165 Friday, July 11, 2014 5:12 PM
Vista de página 180
1 2 ... 176 177 178 179 180 181 182 183 184 185 186 ... 204 205

Comentários a estes Manuais

Sem comentários