Operating the DC Power Analyzer 3
Model N6705A User’s Guide 41
When the Protection function trips:
If an over-voltage, over-current, over-temperature, inhibit signal, a
power-fail condition, or on some modules, a power-limit condition
occurs, the DC Power Analyzer disables the affected output.
To clear the protection function, first remove that condition that
caused the protection fault. Then press the Settings key to access the
Source Settings window. Navigate to and select Protection, then
select Clear All Outputs. This clears the protection function and
returns the output to its previous operating state.
Configure a Turn-On/Turn-Off Sequence
Turn-on and turn-off delays control the turn-on and turn-off
sequencing of the outputs in relation to each other. Press the Settings
key twice to access the Output On/Off Delays window.
Enter the On Delays and Off Delays in milliseconds. Values can
range from 0 milliseconds to 1023 milliseconds in 1-millisecond
increments.
Once output delays have been set, use the All Outputs
key to
start the On delay sequence. Use the All Outputs key to start
the Off delay sequence.
All power modules have an internal delay that applies from the time
that the command to turn on the output is received until the output
actually turns on. This turn on delay is automatically added to the On
delay values that you specify in the Output On/Off Delays window.
Click on Output Coupling to find out what the maximum delay offset
is for the power modules that are installed in the DC Power Analyzer.
This window lets you further customize the operation of the Output
On/Off Delays. Refer to Appendix D for further information.