Agilent N6705A DC Power Analyzer
User’s Guide
A 141
Appendix D
Output On/Off Synchronization
Output Coupling...............................................................................................142
Coupling Multiple Mainframes .....................................................................143
Normally, all outputs in an Agilent N6705A mainframe are included
in an output on/off delay sequence. Additionally, a delay offset is
automatically calculated and applied by the firmware to synchronize
the output turn-on delays.
Output on/off synchronization lets you manually select specific
outputs to be synchronized as well as specify the delay offset, which
serves as a reference for the user-programmed turn-on delays.
This makes it possible have some outputs excluded from an output
on/off delay sequence and be available for other purposes. It also
makes it possible to connect multiple Agilent N6705A mainframes
together and program accurate turn-on delay sequences across
multiple mainframes. Manually specifying a delay offset lets you
configure shorter or longer offset delays than the delay offset that is
automatically applied by the firmware.
NOTE
There is no need to specify a delay offset when outputs turn off. Outputs start
executing their turn-off delays as soon as an output Off command is received.