InfiniiMax Active Probes and Accessories
1130A-34A InfiniiMax high-performance active probe system
1.5 - 7 GHz InfiniiMax probe system
InfiniiMax probe amplifier supports
both differential- and single-ended
measurements for a more cost-
effective solution
Unrivaled InfiniiMax probing
accessories support browsing,
solder-in, and socketed use models
at the maximum performance
available
Compatible with 3000 X-,
4000 X-, 6000 X-, 5000, 6000, and
7000 Series oscilloscopes (except
for 6000 Series 100-MHz models)
The InfiniiMax 1130A Series probe
amplifier is a perfect complement
to the InfiniiVision 1 GHz or higher
bandwidth models. Its bandwidth,
extremely low input capacitance
(0.32 pF), high common mode
rejection, and patented resistor
probe tip technology provide ultra
low loading of the DUT and superior
signal fidelity. Agilent’s innovative
InfiniiMax 1130A Series differential
probe is the easiest-to-use, and
highest performance probing system
available for high-speed digital
design, and represents a new industry
standard for accuracy, flexibility, and
reliability. Designers can achieve full
system bandwidth in conjunction
with InfiniiVision Series oscilloscopes
even when manually browsing
with the probe or making hands-off
measurements. Optional solder-in
probe heads and solder-in sockets, as
well as browser configuration provide
full bandwidth at the probe tip.
Operating characteristics
Probe bandwidth (–3 dB) 1134A: > 7 GHz
1132A: > 5 GHz
1131A: > 3.5 GHz
1130A: > 1.5 GHz
Rise and fall time
(10 to 90%)
1134A: 60 psec
1132A: 86 psec
1131A: 100 psec
1130A: 233 psec
Input capacitance Cm = 0.1 pF Cm is between tips
Cg = 0.34 pF Cg is ground for each tip
Cdiff = 0.27 pF Differential mode capacitance = Cm + Cg/2
Cse = 0.44 pF Single-ended mode capacitance = Cm + Cg
Input resistance Differential mode resistance = 50 k ± 2%
Single-ended mode resistance = 25 k ± 2%
Input dynamic range ±2.5 V
Input common mode range ±6.75 Vdc to 100 Hz; ±1.25 V > 100 Hz
Maximum signal slew rate 18 V/ns when probing a single-ended signal
30 V/ns when probing a differential signal
DC attenuation 10:1 ± 3% before calibration on oscilloscope
10:1 ± 1% after calibration on oscilloscope
Offset range ±12.0 V when probing single-ended signal
Maximum input voltage 30 Vpeak, CAT I
ESD tolerance > 8 kV from 100 pF, 300  HBM
Maximum number of
probes supported by
3000 X-/5000/6000/7000
Series
2
Maximum number of
probes supported by
4000 X- and 6000 X-Series
4
Agilent 1130A InfiniiMax probe offers you the highest performance available for
measuring differential and single-ended signals
Vista de página 6