10
Characteristics and specifications for N2750A Series InfiniiMode probes
Model number N2750A N2751A N2752A
Probe bandwidth* (–3 dB) 1.5 GHz 3.5 GHz 6 GHz
Rise time, probe only (10-90%) 233 psec 100 psec 58.3 psec
Input resistance (@DC)* 200k ± 2% (differential mode)
100k ± 2% (single-ended mode)
50k ± 2% (common mode)
Input capacitance 700 fF (with browser)
Attenuation ratio (@DC) 2:1 / 10:1
Input dynamic range ±1 V, 2 Vpp (@2:1) / ±5 V, 10 Vpp (@10:1)
Input common mode range ±15 V (from DC to 100 Hz), ±2.5 V (for >100 Hz) ***
Offset range ±15 V
Offset accuracy** < 3%
Maximum non-destructive
Input Voltage
±30 V (DC + peak AC)
Maximum number of probes supported
by 3000 X-Series
1
Maximum number of probes supported
by 4000 X-Series
4
* Denotes warranted electrical specifications at 2:1 attenuation mode after 20 minute warm-up, All others are typical.
** When calibrated on the oscilloscope, these characteristics are determined by the oscilloscope characteristics.
*** Assumes symmetric differential signals.
N2750A with browser tip N2750A with socketed tip N2750A with solder-in tip
For more information about the N2750A Series InfiniiMode probes, refer to the data sheet with the Agilent literature
number, 5991-0560EN.