336 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:MEASure:XMIN
(see page 750)
Command Syntax
:MEASure:XMIN [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:XMIN command installs a screen measurement and starts
an X-at- Min-Y measurement on the selected window. If the optional source
parameter is specified, the current source is modified.
Query Syntax
:MEASure:XMIN? [<source>]
The :MEASure:XMIN? query measures and returns the horizontal axis
value at which the minimum vertical value occurs. If the optional source is
specified, the current source is modified. If all channels are off, the query
returns 9.9E+37.
Return Format
<value><NL>
<value> ::= horizontal value of the minimum in NR3 format
See Also "Introduction to :MEASure Commands" on page 290
":MEASure:XMAX" on page 335
":MEASure:TMIN" on page 684
NOTE
:MEASure:XMIN is an alias for :MEASure:TMIN.