Agilent Technologies N9340A Guia do Utilizador Página 150

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 179
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 149
6 SCPI Command Reference
144 N9340A User’s Guide
Types of Detection
[:SENSe]:DETector:TRACe[1]|2|3|4
[FUNCtion]
AVERage|POSitive|SAMPle| NEGative|NORMal
[:SENSe]:DETector:TRACe[1]|2|3|4
[FUNCtion]?
This command specifies the detection mode. For
each trace interval (bucket), average detection
displays the average of all the samples within the
interval. The averaging can be done using two
methods:
The power method (RMS)
The video method (Y Axis Units)
The method is controlled by the TRACE, Detector.
FREQuency Subsection
Center Frequency
[:SENSe]:FREQuency:CENTer <freq>
[:SENSe]:FREQuency:CENTer UP|DOWN
[:SENSe]:FREQuency:CENTer?
This command sets the center frequency.
Center Frequency Step Size Automatic
[:SENSe]:FREQuency:CENTer:STEP:AUTO
OFF|ON|0|1
[:SENSe]:FREQuency:CENTer:STEP:AUTO?
This command specifies whether the step size is
set automatically based on the span.
*RST Value: 1.5 GHz
Available unit: Hz, kHz, MHz, GHz
Key Access: [FREQ] > {Center Freq}
*RST State: On
Key Access: [FREQ] > {CF Step}