Agilent Technologies N9340A Guia do Utilizador Página 158

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 179
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 157
6 SCPI Command Reference
152 N9340A User’s Guide
Amplitude Step
:TGENerator:AMPLitude:STEP <ampt>
:TGENerator:AMPLitude:STEP?
This command sets the amplitude step size of the
tracking generator output. This command is valid
after the tracking generator is enabled.
Amplitude Offset
:TGENerator:AMPLitude:OFFSet <ampt>
This command sets the amplitude offset of the
tracking generator output. This command is valid
after the tracking generator is enabled.
Storing as reference
:TGENerator:NORMalize:REF
This command stores the current trace to Trace 4
as a reference. This command is valid after the
tracking generator is enabled.
*RST State: 1 dB
Range: 1 to 10 dB
Min. Increment 1 dB
Available unit: dB
Key Access: {Amptd Step}
*RST State: 0.00 dB
Range: –88.40 to 327.60 dB
Min. Increment 0.01 dB
Available unit: dB
Key Access: {Amptd Offset}
Key Access: {Normalize} > {Store Ref}
Vista de página 157
1 2 ... 153 154 155 156 157 158 159 160 161 162 163 ... 178 179

Comentários a estes Manuais

Sem comentários