The Agilent N6705 DC power ana-
lyzer can also function as a datalog-
ger. Using the measurement capa-
bility built into each DC power mod-
ule, the N6705 can continuously log
data to the large color display and
to a file. Data can be logged all four
DC outputs at the same time. The
accuracy of the logged voltage and
current measurements are based on
the type of module that is installed
(basic, high-performance, preci-
sion, and SMU).
There are two modes of operation:
In standard mode, measurements
are made spaced apart by the
sample period, which is program-
mable from 75 milliseconds to
60 seconds. For each DC output,
the logged measurements can be
voltage measurements, current
measurements, or both. Each
reading is an integrated voltage or
current measurement. Standard
mode datalogging is available on
all DC module types.
In continuously sampled mode,
the built-in digitizer of the DC
power module runs continuously
at 50,000 readings per second.
You can specify a sample period,
which is the period of time during
which these continuous readings
will be accumulated. For each
sample period, one average read-
ing (and option
ally, a minimum and
maximum
value) will be saved.
In this mode, the digitizer runs
continuously as the readings are
averaged and stored; therefore,
the digitizer is always making
measurements and no data is
missed. The sample period is
programmable from 1 millisecond
to 60 seconds. In this mode, the
N6760 Series of precision DC
power modules offer simultane-
ous logging of output voltage and
output current. For all other mod-
ule types, you can select to log
either voltage or a current when
in continuous sampling mode.
The maximum datalog file size
is 2 gigabytes, which is approxi-
mately 500 million readings. The
logged data file can be stored on
the N6705’s internal non-volatile
RAM or saved externally on a USB
memory device.
The datalogger display can be saved
as a GIF file for use in reports. The
logged data can be saved for view-
ing at a later time. Logged data can
also be exported to a CSV file that
can be read by most common data
analysis software packages.
6
Datalogger
Figure 6. In Datalog View, you can log
data on multiple traces. Here, the current
flowing out of output 1 and output 2 are
captured over 30 seconds
Figure 7. To set up the datalogger,
you use a menu screen to select the
operating parameters. Menu screens like
this are used throughout the DC Power
Analyzer for setup