17
Agilent N6731B – N6736B and N6741B – N6746B Performance Specifications
Unless otherwise noted, specifications are warranted over the ambient temperature range of 0 to 55°C after a 30-minute warm-up period,
with each module’s sense terminals externally jumpered directly to their respective output terminals (local sensing)
N6731B/ N6732B/ N6733B/ N6734B/ N6735B/ N6736B/
N6741B N6742B N6743B N6744B N6745B N6746B
DC output ratings:
Voltage 5 V 8 V 20 V 35 V 60 V 100 V
Current
NOTE 1
10 A / 20 A 6.25 A / 12.5 A
NOTE 3
2.5 A / 5 A 1.5 A / 3 A 0.8 A / 1.6 A 0.5 A / 1 A
Power 50 W / 100 W 50 W / 100 W 50 W / 100 W 52.5 W / 105 W 50 W / 100 W 50 W / 100 W
Output ripple
and noise (PARD)
(from 20 Hz –
20 MHz) CV peak-to- peak 10 mV / 11 mV 12 mV 14 mV 15 mV 25 mV 30 mV
CV rms 2 mV 2 mV 3 mV 5 mV 9 mV 18 mV
Load effect
(Regulation)
(with output change Voltage 5 mV 6 mV 9 mV 11 mV 13 mV / 16 mV 20 mV / 30 mV
from no load to Current 2 mA 2 mA 2 mA 2 mA 2 mA 2 mA
full load, up to a
maximum load-lead
drop of 1 V/lead)
Source effect
(Regulation) Voltage 1 mV 2 mV 2 mV 4 mV 6 mV 10 mV
Current 1 mA 1 mA 1 mA 1 mA 1 mA 1 mA
Programming
accuracy
(@ 23 °C ±5°C after Voltage 0.1% + 19 mV 0.1% + 19 mV 0.1% + 20 mV 0.1% + 35 mV 0.1% + 60 mV 0.1% + 100 mV
30 minute warm-up. Current 0.15% + 20 mA 0.15% + 20 mA 0.15% + 20 mA 0.15% + 20 mA 0.15% + 20 mA 0.15% + 10 mA
Applies from minimum
to maximum program-
ming range)
Voltmeter/ammeter
measurement accuracy
(at 23°C ±5°C) Voltage 0.1% + 20 mV 0.1% + 20 mV 0.1% + 20 mV 0.1% + 35 mV 0.1% + 60 mV 0.1% + 100 mV
Current 0.15% + 20 mA 0.15% + 10 mA 0.15% + 5 mA 0.15% + 4 mA 0.15% + 4 mA 0.15% + 2 mA
Load transient
recovery time
(time to recover to within the settling band following a load
change from 50% to 100% and from 100% to 50% of full load.)
Voltage settling band
±0.08 V / 0.1 V
NOTE 2
±0.08 V / 0.1 V
NOTE 2
± 0.2 V / 0.3 V ± 0.2 V / 0.3 V ± 0.4 V / 0.5 V ± 0.5 V / 1.0 V
Time < 200 µs < 200 µs < 200 µs < 200 µs < 200 µs < 200 µs
1
Output current is derated 1% per °C above 40°C.
2
Settling band is ±0.10 V/0.125 V for 5 V and 8 V Models when relay options 760 and 761 are installed.
3
For N6742B, output current is limited to 10 A when option 760 Output Disconnect/Polarity Reversal Relays is installed.