Power Meter Operation
Setting Measurement Limits
2-48 Agilent E4418B Users Guide
Figure 2-14: Pass/Fail Limit Indicators
This measurement
has passed. This is
indicated by the
limit field being
empty.
This measurement
has failed as the
result is less than
the minimum limit
level set of
-55 dBm. This is
indicated with the
text Undr Lmt.
This
measurement has
failed as the result
is more than the
maximum limit
level set of
-60 dBm. This is
indicated with the
text Over Lmt.