Power Meter Operation
Recorder Output
2-60 Agilent E4418B Users Guide
Leveling a Source Output
The Recorder output can be used to level an output from a source using
external leveling. The following procedure explains how to do this:
1. Select the Recorder Output by pressing , ,
.
2. The highest power you are going to measure is used to determine
the value which you should set for the Recorder Output maximum
setting. For example, if you are measuring a power less than 1 mW
and greater than 100 µW, then set the recorder maximum value to
1mW.
To set the maximum value press and enter the
appropriate value.
3. Press and enter 0 W.
4. Press to On.
50 dBm (100 W)
40 dBm (10 W)
30 dBm (1 W)
20 dBm (100 mW)
10 dBm (10 mW)
0dBm (1mW)
-10 dBm (100 µW)
-20 dBm (10 µW)
-30 dBm (1 µW)
-40 dBm (100 nW)
-50 dBm (10 nW)
-60 dBm (1 nW)
System
Inputs
More
Recorder Output
Max Power
Min Power
Output Off On
HP4402.book Page 60 Wednesday, November 29, 2000 4:46 PM