Power Meter Operation
Making Measurements with the Agilent E-Series Power Sensors
2-16 Agilent E4418B Users Guide
Example
To make a measurement using an Agilent E-series power sensor. The
frequency of the signal to be measured is 100 MHz.
Disconnect the power sensor from any source.
Press .
Press .
Connect the power sensor to the POWER REF output.
Press .
Press , . Use the , , and
hardkeys to enter 100. Press .
Connect the power sensor to the signal to be measured.
The measurement result is now displayed.
Zero
Cal
Zero
Cal
Frequency
Cal Fac
Freq
MHz
HP4402.book Page 16 Wednesday, November 29, 2000 4:46 PM