Power Meter Operation
Setting Measurement Limits
2-44 Agilent E4418B Users Guide
Figure 2-11: Limits Checking Application
Figure 2-12: Limits Checking Results
In this application a swept frequency signal is applied to the input of the
Device Under Test. The power meter measures the output power. The
limits have been set at +4 dBm and +10 dBm. A fail occurs each time the
output power is outside these limits as shown in Figure 2-12.
Setting Window Limits
The power meter can be configured to verify the current measurement in
either window against predefined upper and/or lower limit values. The
range of values that can be set for the upper and lower limits and the
Power Meter
Swept Source
CHANNEL A
INPUT
OUT
IN
OUT
Device
Under Test
4 dBm
+
10 dBm
Amplitude
Frequency
o
o
o
o
o
o
o
Fail
Fail
HP4402.book Page 44 Wednesday, November 29, 2000 4:46 PM