3Capturing Data
86 Agilent 1000 Series Oscilloscopes User’s Guide
5 Press Source and continue pressing the softkey or turn the entry
knob to select the waveform to trigger on:
6 Press Slope and continue pressing the softkey or turn the entry knob
to select the edge to trigger on:
To set up pulse width triggers
A pulse width trigger occurs when a pulse that matches the pulse
definition is found in a waveform.
The width setting can be adjusted from 20 ns to 10 s.
1 Press [Menu].
2 In the Trigger menu, press Mode.
3 Continue pressing the Mode softkey or turn the entry knob to select
“Pulse”.
4 Then, either push the entry knob or press Mode again.
CH1 - CH4 The oscilloscope input channel.
EXT The external trigger input.
EXT/5 The (5:1) attenuated external trigger input.
AC Line The AC power line.
Rising edge.
Falling edge.
Both rising and falling edges.