Making Measurements 4
Agilent 1000 Series Oscilloscopes User’s Guide 107
Phase Between Rising Edges
Measures the phase between two waveforms using the rising edges.
Phase is the calculated phase shift from source 1 to source 2, expressed in
degrees. Negative phase shift values indicate that the rising edge of source
1 occurred after the rising edge of source 2.
Phase Between Falling Edges
Measures the phase between two waveforms using the falling edges. See
Figure 45 on page 107.
Figure 45 Phase Measurements
Phase
Delay
Source 1 Period
--------------------------------------
360=
Phase from channel 1
to channel 2 rising edge
Time
origin
line
50%
Channel 2
Channel 1
50%
Delay
Period