Agilent Technologies 1000A Guia do Utilizador Página 22

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 158
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 21
1Getting Started
22 Agilent 1000 Series Oscilloscopes User’s Guide
Step 5. Use Auto-Scale
The oscilloscope has an auto- scale feature that automatically sets the
oscilloscope controls for the input waveforms present.
Auto-scale requires waveforms with a frequency greater than or equal to
50 Hz and a duty cycle greater than 1%.
1 Press the front panel [Auto-Scale] key.
2 When the AUTO menu appears, press [Menu On/Off] to turn off the
menu.
The oscilloscope turns on all channels that have waveforms applied and
sets the vertical and horizontal scales appropriately. It also selects a
time base range based on the trigger source. The trigger source selected
is the highest- numbered channel that has a waveform applied.
(The Undo softkey in the AUTO menu lets you cancel the auto-scale and
go back to the previous setup.)
Figure 3 [Auto-Scale] Key
Menu
Zoom
Cursors
Measure
Acquire
Display
Help
Utility
Print
Save
Recall
Default
Setup
Auto-
Scale
Run
Stop
Single
Push
Push
Push
for Zoom
to Zero
for 50%
Vista de página 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 157 158

Comentários a estes Manuais

Sem comentários