29
U2500A
Test condition
Dynamic range test for U2500A Series DAQ devices
Dynamic range test Model Test conditions (DUT setting at ±10 V bipolar)
SFDR, THD, SINAD, SNR, ENOB
U2531A
Sampling rate:
Fundamental frequency:
Number of points:
Fundamental input voltage:
2 MSa/s
19.927 kHz
65536
FSR –1 dB FS
U2541A
Sampling rate:
Fundamental frequency:
Number of points:
Fundamental input voltage:
250 kSa/s
2.4109 kHz
8192
FSR – 1 dBFS
U2542A
Sampling rate:
Fundamental frequency:
Number of points:
Fundamental input voltage:
500 kSa/s
4.974 kHz
16384
FSR – 1 dBFS
Bandwidth test for U2500A Series DAQ devices
Bandwidth test Model Test conditions (DUT setting at ±10 V bipolar)
–3 dB Small signal bandwidth:
1% THD Large signal bandwidth:
U2531A
Sampling rate:
Input voltage
–3 dB Small signal bandwidth:
1% THD Large signal bandwidth:
2 MSa/s
10% FSR
FSR – 1 dBFS
U2541A
Sampling rate:
Input voltage
–3 dB Small signal bandwidth:
1% THD Large signal bandwidth:
250 kSa/s
10% FSR
FSR – 1 dBFS
U2542A
Sampling rate:
Input voltage
–3 dB Small signal bandwidth:
1% THD Large signal bandwidth:
500 kSa/s
10% FSR
FSR – 1 dBFS