Agilent Technologies N5230C Guia do Utilizador Página 84

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 102
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 83
76 User’s Guide
U3042AE12 Operational Check
Figure 60 Calibration Complete
6. Repeat step 1 thru step 5 (1-Port Calibration Procedure) for Ports 2 thru 16. When
finished, there should be sixteen Cal Sets saved with the titles “999.1” thru “999.16”
(16-Port).
If you are using an ECal module you can verify the individual port calibration by
selecting Response > CAL > More > ECAL > ECAL Confidence Check. For further
information refer to the system Help menu.
7. Press Trace/Chan > Trace > Delete Trace. There should be no traces on the PNA screen.
Vista de página 83
1 2 ... 79 80 81 82 83 84 85 86 87 88 89 ... 101 102

Comentários a estes Manuais

Sem comentários