Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 3- 35
Application Test Definition
GPIB I/O
Read List Reads a vector data from the specified GPIB device.
Address GPIB address of the target device
Values Vector variable name. Used to store the returned value.
Delimiter Delimiter. 1 character. 0 to 9, E, e, +, -, and . (comma) cannot
be specified.
Length Actually
Read
Numeric variable name. Used to store the number of data stored
in the vector variable.
Terminator Terminator, DEFAULT, EOI, CR/LF+EOI, LF+EOI, CR/LF,
LF, or NONE (not specified)
Timeout Timeout. The value must be less than 2
31
s, 1 ms resolution.
0
does not set timeout.
Read STB Reads the status byte of the specified GPIB device.
Address GPIB address of the target device
Status Byte Numeric variable name. Used to store the status byte to read.
Timeout Timeout. The value must be less than 2
31
s, 1 ms resolution.
0
does not set timeout.
GP-IB Control This area selects one from the following GPIB statements to execute.
Group Execution Trigger
Sends a trigger to the specified GPIB device.
Address GPIB address of the target device
Timeout Timeout. The value must be less than 2
31
s, 1 ms resolution.
0
does not set timeout.
Device Clear
Sends a device clear to the specified GPIB device.
Address GPIB address of the target device
Timeout Timeout. The value must be less than 2
31
s, 1 ms resolution.
0
does not set timeout.