Agilent Technologies E1401-90493 Manual do Utilizador Página 391

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 642
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 390
Chapter 2 391
Performance Verification Tests
53. IF INPUT Accuracy (Option AYZ only)
Figure 2-73 IF INPUT Accuracy Test Setup
Procedure
Determining the IF Centering Error
1. Preset the synthesized sweeper and set the controls as follows:
CW321.4 MHz
POWER LEVEL–30 dBm
RFOn
2. Connect the synthesized sweeper output to the analyzer
50
Input connector. The analyzer provides the frequency reference
for the synthesized sweeper.
3. Press
Preset on the analyzer. Press the Factory Preset softkey, if it is
displayed. Set the analyzer by pressing the following keys:
System, Alignments, Auto Align, Off
FREQUENCY
, 321.4 MHz
SPAN, 5 kHz
AMPLITUDE, –30 dBm
BW/Avg, Res BW, 1 kHz
4. On the analyzer, press Single and wait for the sweep to finish.
5. On the analyzer, press the following keys:
Peak Search (or Search)
Marker, Delta
System
, Alignments, Freq Correct (Off).
Vista de página 390
1 2 ... 386 387 388 389 390 391 392 393 394 395 396 ... 641 642

Comentários a estes Manuais

Sem comentários