Agilent Technologies E1401-90493 Manual do Utilizador Página 211

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 642
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 210
Chapter 2 211
Performance Verification Tests
24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B
MHz (ac) column of Table 2-50. Record this number in Table 2-56 as
the Minimum Response for Band 0.
3. For Band 0 in Table 2-56, subtract the Minimum Response value
from the Maximum Response value and record the result in the
Peak-to-Peak Response column.
4. Note the most positive number from the Flatness Relative to 50 MHz
(ac) column of Table 2-50 for frequencies between 800 MHz and
1.0 GHz. Record this number in Table 2-56 as the Maximum
Response for Band 0A.
5. Note the most negative number from the Flatness Relative to
50 MHz (ac) column of Table 2-50 for frequencies between 800 MHz
and 1.0 GHz. Record this number in Table 2-56 as the Minimum
Response for Band 0A.
6. For Band 0A in Table 2-56, subtract the Minimum Response value
from the Maximum Response value and record the result in the
Peak-to-Peak Response column.
7. Note the most positive number from the Flatness Relative to 50 MHz
(ac) column of Table 2-50 for frequencies between 1.7 GHz and
2.0 GHz. Record this number in Table 2-56 as the Maximum
Response for Band 0B.
8. Note the most negative number from the Flatness Relative to
50 MHz (ac) column of Table 2-50 for frequencies between 1.7 GHz
and 2.0 GHz. Record this number in Table 2-56 as the Minimum
Response for Band 0B.
9. For Band 0B in Table 2-56, subtract the Minimum Response value
from the Maximum Response value and record the result in the
Peak-to-Peak Response column.
10.Note the most positive number from the Flatness Relative to 50 MHz
(ac) column of Table 2-52 for frequencies between 3.0 GHz and
6.7 GHz. Record this number in Table 2-56 as the Maximum
Response for Band 1.
11.Note the most negative number from the Flatness Relative to
50 MHz (ac) column of Table 2-52 for frequencies between 3.0 GHz
and 6.7 GHz. Record this number in Table 2-56 as the Minimum
Response for Band 1.
12.For Band 1 in Table 2-56, subtract the Minimum Response value
from the Maximum Response value and record the result in the
Peak-to-Peak Response column.
13.Note the most positive number from the Flatness Relative to 50 MHz
(ac) column of Table 2-52 for frequencies between 6.7 GHz and
13.2 GHz. Record this number in Table 2-56 as the Maximum
Response for Band 2.
Vista de página 210
1 2 ... 206 207 208 209 210 211 212 213 214 215 216 ... 641 642

Comentários a estes Manuais

Sem comentários