Chapter 2 247
Performance Verification Tests
28. Other Input-Related Spurious Responses: Agilent E4401B and E4411B
24.On the synthesized sweeper, press RECALL, 2.
25.On the analyzer, press
Single and wait for the completion of a new
sweep.
26.On the analyzer, press
Peak Search (or Search), record the marker
delta (
Mkr1) amplitude in the performance verification test record
as indicated in Table 2-67.
27.On the synthesized sweeper, press
RECALL, 4.
28.On the analyzer press the following keys:
FREQUENCY, Center Freq, 100 MHz
AMPLITUDE
, –5 dBm(50 Input)
AMPLITUDE, 48.75 dBmV (75 Input)
Marker, Normal
Sweep, Sweep
(Cont)
Peak Search (or Search), Meas Tools, Mkr Ref Lvl
Peak Search
(or Search), Meas Tools, Delta
Single
29.On the synthesized sweeper, press RECALL 3 for a CW frequency of
1310.7 MHz.
30.Press
Single on the analyzer and wait for a completion of a new
sweep.
31.On the analyzer, press
Peak Search (or Search) and record the marker
delta (
Mkr1) amplitude in the performance test record as indicated
in Table 2-67.