2-32
Making Mixer Measurements (Option 089 Only)
Fixed IF Mixer Measurements
Press and the analyzer will
display the following sequence commands:
SEQUENCE SEQ2
Start of Sequence
WAIT x
1 x1
MANUAL TRG ON POINT
TITLE
FREQ:CW UP
PERIPHERAL HPIB ADDR
19x1
TITLE TO PERIPHERAL
PERIPHERAL HPIB ADDR
21x1
TITLE TO PERIPHERAL
DECR LOOP COUNTER
IF LOOP COUNTER <>0 THEN DO
SEQUENCE 2
TITLE
MEASUREMENT COMPLETED
2. Press the following keys to run the sequences:
When the prompt CONNECT MIXER appears, connect the equipment as shown in Figure
2-23.
Figure 2-23 Connections for a Conversion Loss Using Tuned Receiver Mode
Seq
NEW SEQ/MODIFY SEQ
SEQUENCE 2 SEQ2
Seq
DONE SEQ MODIFY
DO SEQUENCE
SEQUENCE 2 SEQ2