Programming Reference
SCPI Instrument Command List
Command :PULS:DOUB[1]:DEL:HOLD
Long
[:SOURce]:PULSe:DOUBle[1]:DELay:HOLD
Form Set & Query
Parameter
TIME|PRATio
*RST value
TIME
Description Use this command to set/read the coupling between the pulse period and
the double-pulse delay:
Example To set and hold the double-pulse delay:
Command :PULS:DOUB[1]:DEL:UNIT
Long
[:SOURce]:PULSe:DOUBle[1]:DELay:UNIT
Form Set & Query
Parameter
S | SEC | PCT
*RST value
S
Description Use this command to set/read the default units for the double delay
parameter. The default unit of a parameter is the unit used when the
parameter is programmed to a value without a unit suffix.
Example To set the double-pulse delay to 50%:
TIME The absolute double-pulse delay is held fixed when the
pulse period is varied.
PRATio The double-pulse delay as percentage of period is held
fixed when the pulse period is varied.
:PULS:DOUB ON
:PULS:DOUB:DEL 50 PCT
:PULS:DOUB:DEL:HOLD PRAT
Switch on double-pulses on OUTPUT
Set interpulse delay to 50% of pulse
period
Hold interpulse delay as fixed percent-
age of pulse period
:PULS:DOUB:DEL:UNIT PCT
:PULS:DOUB:DEL 50
Set OUTPUT double delay unit to %
Set OUTPUT double-pulse delay to 50%
of period