Programming Reference
SCPI Instrument Command List
Command :SYST:WARN:STR?
Long
:SYSTem:WARNing:STRing?
Form Query
*RST value Not applicable
Description Use this command to read all the currently active warning messages. The
warning messages are concatenated to form a single string with a “;” as
separator between the messages.
Command :SYST:WARN:BUFF?
Long
:SYSTem:WARNing:BUFFer?
Form Query
*RST value Not applicable
Description Use this command to read the maximum possible number of characters
that could be returned by
:SYST:WARN:STR?
if all warnings were active.
Command :TRIG:COUNt
Long
:TRIGger[:SEQuence[1]]:COUNt
Form Set & Query
Parameter Numeric
*RST value 1
Specified limits 1 to 65 536
Description Use this command to set/read the number of trigger events (pulse
periods) to be generated for each arming event. This corresponds to
selecting the event mode on the M
ODE
/T
RIGGER
SCREEN
:
PULSES
Set a trigger count to 1 so that a single pulse period is generated for
each arming event. The instrument is in pulse (stream) mode.
BURST of
Set a trigger count of 2 to 65536 so that a burst of 2 to 65536 pulse
periods is generated for each arming event.