Agilent Technologies 222A Manual de Instruções Página 72

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 123
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 71
72
Programming Reference
SCPI Instrument Command List
Command :VOLT[1]:HIGH
Long
[:SOURce]:VOLTage[1][:LEVel][:IMMediate]:HIGH
Form Set & Query
Parameter Numeric
Parameter suffix
V
with engineering prefixes.
Value coupling
Range coupling With low level, see page 73
*RST value 500 mV
Specified limits 9.9 V to 10.0 V (50
into 50
)
Description Use this command to program the high level voltage of the OUTPUT
signal. Note that to set the OUTPUT levels in terms of voltage, you first
have to execute the
[:SOURce]:HOLD VOLTage
command to enable the
[:SOURce]:VOLTage
subsystem.
The available voltage range is limited by the combination of:
Specified current limits
Actual OUTPUT impedance setting
:OUTPut:IMPedance
Actual expected load impedance setting
:OUTPut:IMPedance:EXTernal
Example To set the high level voltage:
Amplitude = High – Low
Offset =
High – Low
2
:HOLD VOLT
:VOLT:HIGH 4.8V
Enable VOLTAGE subsystem
Set OUTPUT high level voltage to 4.8 V
Vista de página 71
1 2 ... 67 68 69 70 71 72 73 74 75 76 77 ... 122 123

Comentários a estes Manuais

Sem comentários