Agilent-technologies E1300B Manual do Utilizador Página 169

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 244
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 168
:COMMunicate
:SERial[n] [:RECeive]
:PACE [:PROTocol]
SYSTem:COMMunicate:SERial[n][:RECeive]:PACE[:PROTocol]
< protocol> enables or disables receive pacing (XON/XOFF) protocol.
Parameters Parameter
Name
Parameter
Type
Range of
Values
Default
Units
protocol discrete XON| NONE none
Comments While PROT is XON, the serial interface will send XOFF when the
buffer reaches the STOP threshold, and XON when the buffer reaches
the STARt threshold.
For an Agilent E1324A, AUTO is always ON. In this case
[:RECeive]:PACE will also set TRAN:PACE
The XON character is control Q (ASCII 17
10
, 11
16
), The XOFF
character is control S (ASCII 19
10
, 13
16
).
DIAG:BOOT:COLD will set PACE to XON.
Related Commands: PROTocol:THReshold:STARt,
PROTocol:THReshold:STOP, TRAN:AUTO
*RST Condition: No change
Example Enabling XON/XOFF handshaking.
SYST:COMM:SER0:PACE:PROT XON
:COMMunicate
:SERial[n] [:RECeive]
:PACE [:PROTocol]?
SYSTem:COMMunicate:SERial[n][:RECeive]:PACE[:PROTocol]? returns the
current receive pacing protocol.
Example See if XON/XOFF protocol is enabled.
SYST:COMM:SER0:PACE:PROT?
enter statement statement enters the string
"X O N " or "N ON E "
SYSTem :COMMunicate :SERial[n] [:RECeive] :PACE [:PROTocol]
System Instrument Command Reference 7-41
Vista de página 168
1 2 ... 164 165 166 167 168 169 170 171 172 173 174 ... 243 244

Comentários a estes Manuais

Sem comentários