Agilent Technologies
Z5623A Option H48
Multiport Test Set
Users and Service Guide
Manufacturing Part Number: Z5623-90016
Printed in USA
June 2005
© Copyright 2000-2002, 2005 Agilent Technologies, Inc. All rights reserved.
Vista de página 0
1 2 3 4 5 6 ... 137 138

Resumo do Conteúdo

Página 1 - Z5623A Option H48

Agilent TechnologiesZ5623A Option H48Multiport Test SetUser’s and Service GuideManufacturing Part Number: Z5623-90016Printed in USAJune 2005© Copyrigh

Página 2 - Warranty Statement

1-2 Agilent Technologies Z5623A Option H48Instrument DescriptionOverviewOverviewThe Agilent Z5623A Option H48 Multiport Test Set is designed for use

Página 3 - Definitions

5 -32 Agilent Technologies Z5623A Option H48Performance VerificationSetting Up Limit TestingFigure 5-35 The Status Bar Displayed at Bottom of Screen

Página 4

Agilent Technologies Z5623A Option H48 5 -33Performance VerificationSetting Up Limit TestingFigure 5-36 The “Scale...” ButtonStep 6. The Scale window

Página 5 - Contents

5 -34 Agilent Technologies Z5623A Option H48Performance VerificationSetting Up Limit TestingFigure 5-37 The Scale Window With Desired Values For Cro

Página 6 - 5. Performance Verification

Agilent Technologies Z5623A Option H48 5 -35Performance VerificationSetting Up Limit TestingTurning On Limit Lines and Limit TestingStep 1. Click Trac

Página 7

5 -36 Agilent Technologies Z5623A Option H48Performance VerificationVerifying Return Loss and Insertion Loss SpecificationsVerifying Return Loss and

Página 8 - 10. Contacting Agilent

Agilent Technologies Z5623A Option H48 5 -37Performance VerificationVerifying Return Loss and Insertion Loss SpecificationsStep 7. Command the test se

Página 9 - Instrument Description

5 -38 Agilent Technologies Z5623A Option H48Performance VerificationVerifying Return Loss and Insertion Loss SpecificationsTable 5-8 Pass/Fail Test

Página 10 - Overview

Agilent Technologies Z5623A Option H48 5 -39Performance VerificationVerifying Crosstalk SpecsVerifying Crosstalk SpecsCrosstalk need only be measured

Página 11

5 -40 Agilent Technologies Z5623A Option H48Performance VerificationVerifying Crosstalk SpecsStep 9. Move the shorts to the next pair of test ports.

Página 12 - Guaranteed Performance

Agilent Technologies Z5623A Option H48 6 -16 Measuring Multiport Devices

Página 13 - Typical Performance

Agilent Technologies Z5623A Option H48 1-3Instrument DescriptionSpecificationsSpecificationsAgilent provides two different types of specifications for

Página 14 - Environmental Requirements

6 -2 Agilent Technologies Z5623A Option H48Measuring Multiport DevicesCalibrating the Test SystemCalibrating the Test SystemAfter the test set has wa

Página 15 - Available Options

Agilent Technologies Z5623A Option H48 6 -3Measuring Multiport DevicesCalibrating the Test SystemFigure 6-1 Calibrating the Test SystemFor the best ac

Página 16 - Shipping

6 -4 Agilent Technologies Z5623A Option H48Measuring Multiport DevicesCalibrating the Test System

Página 17 - Installation

Agilent Technologies Z5623A Option H48 7 -17 Advanced TopicsThis chapter contains the following sections:• Using an External Computer to Control the T

Página 18 - Checking the Shipment

7 -2 Agilent Technologies Z5623A Option H48Advanced TopicsUsing an External Computer to Control the Test SetUsing an External Computer to Control the

Página 19

Agilent Technologies Z5623A Option H48 7 -3Advanced TopicsUsing an External Computer to Control the Test SetBox IdentificationTo read the Box Identifi

Página 20 - Electrical Preparations

7 -4 Agilent Technologies Z5623A Option H48Advanced TopicsUsing an External Computer to Control the Test SetWrite Commands:info$=“refl_01” ‘command f

Página 21

Agilent Technologies Z5623A Option H48 7 -5Advanced TopicsUsing an External Computer to Control the Test SetUsing National Instruments VISAIf you are

Página 22 - Figure 2-2 Power Cables

7 -6 Agilent Technologies Z5623A Option H48Advanced TopicsUsing the Control Lines ConnectorUsing the Control Lines ConnectorThe Control Lines connect

Página 23 - Environmental Preparations

Agilent Technologies Z5623A Option H48 7 -7Advanced TopicsUsing the Control Lines ConnectorFigure 7-1 Control LinesSuggested External CircuitryFigure

Página 24

1-4 Agilent Technologies Z5623A Option H48Instrument DescriptionSpecificationsGuaranteed PerformanceTable 1-1 Agilent Z5623A Option H48 Guaranteed P

Página 25 - Test Set Familiarization

7 -8 Agilent Technologies Z5623A Option H48Advanced TopicsUsing the Control Lines ConnectorFigure 7-2 Switch Driver Circuit For Use With Control Lin

Página 26

Agilent Technologies Z5623A Option H48 7 -9Advanced TopicsUsing the Control Lines ConnectorThe DUT may require a control voltage that is different fro

Página 27 - Rear Panel

7 -10 Agilent Technologies Z5623A Option H48Advanced TopicsUsing the Control Lines Connector

Página 28 - The Line Fuse

Agilent Technologies Z5623A Option H48 8 -18 ServiceThis chapter contains the following sections:• Adjustments• Theory of Operation• Troubleshooting•

Página 29 - ADDRESS:

8 -2 Agilent Technologies Z5623A Option H48ServiceAdjustmentsAdjustmentsThere are no adjustments for the test set.Figure 8-1 Z5623A Option H48 Block

Página 30

Agilent Technologies Z5623A Option H48 8 -3ServiceTheory of OperationTheory of OperationThe theory of operation begins with a general description of t

Página 31

8 -4 Agilent Technologies Z5623A Option H48ServiceTheory of OperationA2 Controller and A3 Switch Driver Board TheoryThe mother and daughter boards pr

Página 32

Agilent Technologies Z5623A Option H48 8 -5ServiceTroubleshootingTroubleshootingThis section contains procedures for troubleshooting the test set to a

Página 33 - Control the Test Set

8 -6 Agilent Technologies Z5623A Option H48ServiceTroubleshootingTroubleshooting the Front Panel BoardTurn the instrument power on and check the foll

Página 34

Agilent Technologies Z5623A Option H48 8 -7ServiceList of Replaceable PartsList of Replaceable PartsThe following table contains the list of replaceab

Página 35 - Click on

Agilent Technologies Z5623A Option H48 1-5Instrument DescriptionSpecificationsTypical PerformanceTable 1-2 Agilent Z5623A Option H48 Typical Performa

Página 36

8 -8 Agilent Technologies Z5623A Option H48ServiceList of Replaceable PartsNOTEThe above parts are unique to this special option. To order replacemen

Página 37

Agilent Technologies Z5623A Option H48 9-19 Safety and Regulatory Information

Página 38 - C:\Progam Files\National

9-2 Agilent Technologies Z5623A Option H48Safety and Regulatory InformationSafety InformationSafety InformationThis document contains two types of sa

Página 39 - How to Send a Manual Command

Agilent Technologies Z5623A Option H48 9-3Safety and Regulatory InformationSafety InformationWARNINGThese servicing instructions are for use by qualif

Página 40

9-4 Agilent Technologies Z5623A Option H48Safety and Regulatory InformationSafety InformationInstrument MarkingsWhen you see this symbol on your inst

Página 41 - A List of GPIB Commands

Agilent Technologies Z5623A Option H48 9-5Safety and Regulatory InformationRegulatory InformationRegulatory InformationStatement of Compliance with IE

Página 42

9-6 Agilent Technologies Z5623A Option H48Safety and Regulatory InformationRegulatory Information

Página 43

Agilent Technologies Z5623A Option H48 10 -110 Contacting Agilent

Página 44 - Configuration

10 -2 Agilent Technologies Z5623A Option H48Contacting AgilentContacting AgilentContacting AgilentBy internet, phone, or fax, get assistance with all

Página 45 - Disaster Recovery

1-6 Agilent Technologies Z5623A Option H48Instrument DescriptionElectrical RequirementsElectrical RequirementsThe alternating-current (AC) power that

Página 46 - No applications running

Agilent Technologies Z5623A Option H48 1-7Instrument DescriptionGeneral CharacteristicsGeneral CharacteristicsWeightNet: Approximately 9 kgShipping: A

Página 47

1-8 Agilent Technologies Z5623A Option H48Instrument DescriptionCleaning and Shipping InstructionsCleaning and Shipping InstructionsCleaningClean the

Página 48

Agilent Technologies Z5623A Option H48 2 -12 InstallationThis chapter contains the following sections:• Checking the Shipment• Recommended Additional

Página 49

2 -2 Agilent Technologies Z5623A Option H48InstallationChecking the ShipmentChecking the ShipmentAfter the test set has been unpacked, keep the origi

Página 50

Agilent Technologies Z5623A Option H48 2 -3InstallationRecommended Additional EquipmentRecommended Additional EquipmentThe equipment listed in Table 2

Página 51 - Analyzer

ii Warranty StatementTHE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITION

Página 52 - Performing the Calibration

2 -4 Agilent Technologies Z5623A Option H48InstallationFamiliarization with Safety RequirementsFamiliarization with Safety RequirementsThis document

Página 53

Agilent Technologies Z5623A Option H48 2 -5InstallationElectrical PreparationsFigure 2-1 Protective Earth GroundWARNINGThis is a Safety Class I produ

Página 54

2 -6 Agilent Technologies Z5623A Option H48InstallationElectrical PreparationsFigure 2-2 Power CablesLNE250V250VLNE250VELNNLE250V125VEENNLLEENNLL220

Página 55

Agilent Technologies Z5623A Option H48 2 -7InstallationEnvironmental PreparationsEnvironmental Preparations1. Ensure that the “Environmental Requireme

Página 56

2 -8 Agilent Technologies Z5623A Option H48InstallationEnvironmental Preparations3. Set up a static safe workstation such as the one shown in Figure

Página 57 - Renaming the Calibration File

Agilent Technologies Z5623A Option H48 2 -9InstallationTest Set FamiliarizationTest Set FamiliarizationThis section familiarizes the user with various

Página 58 - Files” icon

2 -10 Agilent Technologies Z5623A Option H48InstallationTest Set FamiliarizationThe PORT CONNECTION Status LCDThe PORT CONNECTION Status LCD displays

Página 59

Agilent Technologies Z5623A Option H48 2 -11InstallationTest Set FamiliarizationRear PanelFigure 2-6 Rear Panel FeaturesGPIB ConnectorThis connector

Página 60

2 -12 Agilent Technologies Z5623A Option H48InstallationTest Set FamiliarizationThe Line FuseThe line fuse (F 3 A/250 V, part number 2110-0780) and a

Página 61

Agilent Technologies Z5623A Option H48 2 -13InstallationSetting the GPIB Address of the Test SetSetting the GPIB Address of the Test SetA bank of five

Página 62

iiiSafety Notes The following safety notes are used throughout this document. Familiarize yourself with each of these notes and its meaning before p

Página 63

2 -14 Agilent Technologies Z5623A Option H48InstallationConnecting and Turning on the Test SetNOTEThe test set is shipped from the factory with the G

Página 64 - How to Recall a Calibration

Agilent Technologies Z5623A Option H48 2 -15InstallationConnecting and Turning on the Test SetFigure 2-9 Detailed Equipment Connection DiagramNetwork

Página 65

2 -16 Agilent Technologies Z5623A Option H48InstallationConnecting and Turning on the Test Set

Página 66

Agilent Technologies Z5623A Option H48 3-13 Using the Network Analyzer to Control the Test SetThis chapter assumes the following:• The instructions i

Página 67

3-2 Agilent Technologies Z5623A Option H48Using the Network Analyzer to Control the Test SetPreparing the Analyzer to Control the Test SetPreparing t

Página 68

Agilent Technologies Z5623A Option H48 3-3Using the Network Analyzer to Control the Test SetPreparing the Analyzer to Control the Test SetStep 4. In t

Página 69

3-4 Agilent Technologies Z5623A Option H48Using the Network Analyzer to Control the Test SetPreparing the Analyzer to Control the Test SetStep 6. On

Página 70 - Setting Test Limits

Agilent Technologies Z5623A Option H48 3-5Using the Network Analyzer to Control the Test SetPreparing the Analyzer to Control the Test SetStep 9. When

Página 71

3-6 Agilent Technologies Z5623A Option H48Using the Network Analyzer to Control the Test SetPreparing the Analyzer to Control the Test SetStep 12. Us

Página 72 - Test Strategy

Agilent Technologies Z5623A Option H48 3-7Using the Network Analyzer to Control the Test SetHow to Send a Manual CommandHow to Send a Manual CommandTh

Página 74 - Crosstalk

3-8 Agilent Technologies Z5623A Option H48Using the Network Analyzer to Control the Test SetCommon Errors in Manual CommandsNOTEIt is possible to iss

Página 75

Agilent Technologies Z5623A Option H48 3-9Using the Network Analyzer to Control the Test SetA List of GPIB Commands A List of GPIB CommandsNOTEEvery T

Página 76

3-10 Agilent Technologies Z5623A Option H48Using the Network Analyzer to Control the Test SetA List of GPIB CommandsTable 3-3 Control Line CommandsG

Página 77

Agilent Technologies Z5623A Option H48 3-11Using the Network Analyzer to Control the Test SetA List of GPIB CommandsTo connect all test ports to their

Página 78

3-12 Agilent Technologies Z5623A Option H48Using the Network Analyzer to Control the Test SetRestoring the Analyzer to its Normal ConfigurationRestor

Página 79

Agilent Technologies Z5623A Option H48 3-13Using the Network Analyzer to Control the Test SetDisaster RecoveryDisaster RecoveryThis section provides r

Página 80

3-14 Agilent Technologies Z5623A Option H48Using the Network Analyzer to Control the Test SetDisaster RecoveryFigure 3-8 Network Analyzer Screen Wit

Página 81

Agilent Technologies Z5623A Option H48 3-15Using the Network Analyzer to Control the Test SetDisaster RecoveryStep 2. Right-click on the “My Computer”

Página 82

3-16 Agilent Technologies Z5623A Option H48Using the Network Analyzer to Control the Test SetDisaster RecoveryStep 4. Click the Hardware tab shown in

Página 83 - Saving the Limit Test File

Agilent Technologies Z5623A Option H48 3-17Using the Network Analyzer to Control the Test SetDisaster RecoveryFigure 3-13 Device Manager WindowFigure

Página 84

ContentsContents-11. Instrument DescriptionOverview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 85

3-18 Agilent Technologies Z5623A Option H48Using the Network Analyzer to Control the Test SetDisaster RecoveryFigure 3-15 GPIB Device Disabling Mess

Página 86

Agilent Technologies Z5623A Option H48 4-14 Calibrating the Network AnalyzerThis chapter shows you how to calibrate the network analyzer so that it c

Página 87

4-2 Agilent Technologies Z5623A Option H48Calibrating the Network AnalyzerPerforming the CalibrationPerforming the CalibrationStep 1. Make the connec

Página 88

Agilent Technologies Z5623A Option H48 4-3Calibrating the Network AnalyzerPerforming the CalibrationStep 2. Press the Preset button on the Network Ana

Página 89

4-4 Agilent Technologies Z5623A Option H48Calibrating the Network AnalyzerPerforming the CalibrationFigure 4-4 Cal Type Window When It Is First Open

Página 90

Agilent Technologies Z5623A Option H48 4-5Calibrating the Network AnalyzerPerforming the CalibrationFigure 4-7 Electronic Calibration WindowStep 10.

Página 91

4-6 Agilent Technologies Z5623A Option H48Calibrating the Network AnalyzerPerforming the CalibrationOr repeat your cal.”Figure 4-10 Cal Wizard Windo

Página 92

Agilent Technologies Z5623A Option H48 4-7Calibrating the Network AnalyzerRenaming the Calibration FileRenaming the Calibration FileIn this section yo

Página 93

4-8 Agilent Technologies Z5623A Option H48Calibrating the Network AnalyzerRenaming the Calibration FileStep 4. The window shown in Figure 4-12 will p

Página 94 - Limit Testing for Crosstalk

Agilent Technologies Z5623A Option H48 4-9Calibrating the Network AnalyzerRenaming the Calibration FileStep 6. The window shown in Figure 4-14 will po

Página 95 - Power window

Contents-2Malfunctioning Command Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13Malfunctioning Network Analyzer. . . . . . .

Página 96

4-10 Agilent Technologies Z5623A Option H48Calibrating the Network AnalyzerRenaming the Calibration FileStep 8. The window shown in Figure 4-16 will

Página 97

Agilent Technologies Z5623A Option H48 4-11Calibrating the Network AnalyzerRenaming the Calibration FileStep 9. Move the cursor over the filename that

Página 98

4-12 Agilent Technologies Z5623A Option H48Calibrating the Network AnalyzerRenaming the Calibration FileStep 11. Click File on the menu bar, then scr

Página 99

Agilent Technologies Z5623A Option H48 4-13Calibrating the Network AnalyzerRenaming the Calibration FileStep 12. Give the file a new name that include

Página 100 - Setting Up Limit Testing

4-14 Agilent Technologies Z5623A Option H48Calibrating the Network AnalyzerRecalling and Examining CalibrationsRecalling and Examining CalibrationsTh

Página 101

Agilent Technologies Z5623A Option H48 4-15Calibrating the Network AnalyzerRecalling and Examining CalibrationsStep 2. The Recall window shown in Figu

Página 102

4-16 Agilent Technologies Z5623A Option H48Calibrating the Network AnalyzerRecalling and Examining CalibrationsMethod 2. Calibration File Is On “Quic

Página 103

Agilent Technologies Z5623A Option H48 4-17Calibrating the Network AnalyzerRecalling and Examining CalibrationsFigure 4-25 Calibration Turned OnFigur

Página 104 - Specifications

4-18 Agilent Technologies Z5623A Option H48Calibrating the Network AnalyzerRecalling and Examining CalibrationsHow to Examine the Properties of a Cal

Página 105

Agilent Technologies Z5623A Option H48 5 -15 Performance VerificationThis chapter contains the following sections:• Setting Test Limits• Test Strategy

Página 106

ContentsContents-3Limit Testing for Crosstalk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-26Specifying Output Power

Página 107 - Verifying Crosstalk Specs

5 -2 Agilent Technologies Z5623A Option H48Performance VerificationSetting Test LimitsSetting Test LimitsA test limit is the pass/fail criterion for

Página 108

Agilent Technologies Z5623A Option H48 5 -3Performance VerificationSetting Test LimitsApplying Equation 5-1 to the uncertainties listed in Table 5-1 a

Página 109 - Measuring Multiport Devices

5 -4 Agilent Technologies Z5623A Option H48Performance VerificationTest StrategyTest StrategyThe overall test strategy of this chapter is to verify e

Página 110 - Calibrating the Test System

Agilent Technologies Z5623A Option H48 5 -5Performance VerificationTest StrategyInsertion LossFor this measurement, Port 1 of the network analyzer wil

Página 111 - Calibration and Test

5 -6 Agilent Technologies Z5623A Option H48Performance VerificationTest StrategyReturn loss is specified for two different conditions:• Test port act

Página 112

Agilent Technologies Z5623A Option H48 5 -7Performance VerificationSetting Up Limit TestingSetting Up Limit TestingIn this section, you will create th

Página 113 - Advanced Topics

5 -8 Agilent Technologies Z5623A Option H48Performance VerificationSetting Up Limit TestingLimit Testing for Insertion LossRecalling the CalibrationR

Página 114 - Test Set

Agilent Technologies Z5623A Option H48 5 -9Performance VerificationSetting Up Limit TestingStep 2. Click View on the menu bar. When the View menu appe

Página 115 - Languages

5 -10 Agilent Technologies Z5623A Option H48Performance VerificationSetting Up Limit TestingStep 4. Type “Insertion Loss” in the Entry field as shown

Página 116

Agilent Technologies Z5623A Option H48 5 -11Performance VerificationSetting Up Limit TestingStep 6. The Scale window shown in Figure 5-6 will pop up.

Página 117

Contents-4Connector Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4Troubleshooting . . . . . . . . .

Página 118

5 -12 Agilent Technologies Z5623A Option H48Performance VerificationSetting Up Limit TestingSetting Up the Limit TableStep 1. Click Trace on the menu

Página 119 - Suggested External Circuitry

Agilent Technologies Z5623A Option H48 5 -13Performance VerificationSetting Up Limit TestingShortcuts:a. To specify a Segment Type, click the Segment

Página 120

5 -14 Agilent Technologies Z5623A Option H48Performance VerificationSetting Up Limit TestingTurning On Limit Lines and Limit TestingStep 1. Click Tra

Página 121

Agilent Technologies Z5623A Option H48 5 -15Performance VerificationSetting Up Limit TestingFigure 5-13 Insertion Loss Limit Test Being Run On ECal M

Página 122

5 -16 Agilent Technologies Z5623A Option H48Performance VerificationSetting Up Limit TestingStep 2. When the Save As window comes up as shown in Figu

Página 123 - 8 Service

Agilent Technologies Z5623A Option H48 5 -17Performance VerificationSetting Up Limit TestingChecking the Limit Test FileStep 1. Connect an RF cable fr

Página 124 - Adjustments

5 -18 Agilent Technologies Z5623A Option H48Performance VerificationSetting Up Limit TestingStep 4. Recall the limit line file for insertion loss. Th

Página 125 - Theory of Operation

Agilent Technologies Z5623A Option H48 5 -19Performance VerificationSetting Up Limit TestingLimit Testing for Return Loss (Port Active)Recalling the C

Página 126 - Connector Replacement

5 -20 Agilent Technologies Z5623A Option H48Performance VerificationSetting Up Limit TestingStep 5. The Scale window shown in Figure 5-6 will pop up.

Página 127 - Troubleshooting

Agilent Technologies Z5623A Option H48 5 -21Performance VerificationSetting Up Limit TestingFigure 5-21 Completed Limit Table for Return Loss (Port A

Página 128

Agilent Technologies Z5623A Option H48 1-11 Instrument DescriptionThis chapter contains the following sections:• Overview • Specifications• Electrica

Página 129 - List of Replaceable Parts

5 -22 Agilent Technologies Z5623A Option H48Performance VerificationSetting Up Limit TestingSaving the Limit Test FileStep 1. Click File on the menu

Página 130

Agilent Technologies Z5623A Option H48 5 -23Performance VerificationSetting Up Limit TestingStep 4. Recall the limit line file for return loss (port a

Página 131

5 -24 Agilent Technologies Z5623A Option H48Performance VerificationSetting Up Limit TestingLimit Testing for Return Loss (Port Off)The limit test fo

Página 132 - Safety Information

Agilent Technologies Z5623A Option H48 5 -25Performance VerificationSetting Up Limit TestingFigure 5-24 Completed Limit Table for Return Loss (Port O

Página 133 - Cautions

5 -26 Agilent Technologies Z5623A Option H48Performance VerificationSetting Up Limit TestingLimit Testing for Crosstalk Specifying Output PowerStep 1

Página 134 - Instrument Markings

Agilent Technologies Z5623A Option H48 5 -27Performance VerificationSetting Up Limit TestingStep 3. The Power window shown in Figure 5-26 will pop up.

Página 135 - Regulatory Information

5 -28 Agilent Technologies Z5623A Option H48Performance VerificationSetting Up Limit TestingSpecifying IF Bandwidth and AveragingStep 1. Click Sweep

Página 136

Agilent Technologies Z5623A Option H48 5 -29Performance VerificationSetting Up Limit TestingFigure 5-30 IF Bandwidth Window with Value Desired for Cr

Página 137 - Contacting Agilent

5 -30 Agilent Technologies Z5623A Option H48Performance VerificationSetting Up Limit TestingStep 4. When the Average window shown in Figure 5-32 appe

Página 138

Agilent Technologies Z5623A Option H48 5 -31Performance VerificationSetting Up Limit TestingStep 5. Click View on the menu bar. When the View menu app

Comentários a estes Manuais

Sem comentários