4-24 Agilent E5250A Users Guide, Edition 11
Setting up Measurement Environment
Measurement Cable Length
Table 4-6 Guard Capacitances of E5250A Measurement Environment (Typical)
Connection Agilent Model/Part No.
Cable
Length
Guard
Capacitance
from SMU to
E5250A
16494A-003 (for non-Kelvin) 80 cm 75 pF
16494B-003 (for Kelvin) 80 cm 90 pF
16494C-001 (for 4142B) 1.5 m 140 pF
E5250A and
Plug-in Card
E5252A (only 1 card is installed) 145 pF
E5252A (more 1 card) 8 pF/extra card
E5255A (only 1 card is installed) 180 pF
E5255A(more 1 card) 60 pF/extra card
from Card
Output to
Connector
Plate
16494A-001 (for non-Kelvin) 1.5 m 125 pF
16494A-002 (for non-Kelvin) 3 m 240 pF
16494B-001 (for Kelvin) 1.5 m 140 pF
16494B-002 (for Kelvin) 3 m 260 pF
16494C-001 (for 4142B) 1.5 m 140 pF
16494C-002 (for 4142B) 3 m 260 pF
16494D-001 1.5 m 125 pF
16494D-002 3 m 240 pF
16494E-003 3 m 240 pF
from
Connector
Plate to DUT
Agilent Part No. 8121-1191 X m 130 pF per m